Operating the Parvo Medics
Metabolic Measurement Systems
q Gas calibration (TrueMax 2400 system)
q Gas calibration (consolidated system)
q Performing metabolic (VO2) measurements
Return to: Tutorial Home Page
ENS 314 Exercise Physiology Lab Web Site
Note:
Operating instructions of the consolidated system vary from the TrueMax
2400 system only in the gas calibration procedures; all other procedures are
identical. The gas calibration
procedure should be performed prior to each test. However, the flowmeter needs calibration only twice daily.
Gas calibration for the Parvo Medics TrueMax 2400 system (perform prior to each test)
·
From main menu, click on
Gas Calibration
·
Click OK
·
Turn black plastic valve
wrench on calibration gas cylinder (on right front corner of cart)
counterclockwise 90°, then click OK
·
Allow gas calibration
procedure to finish, and when prompted, click OK
·
Return valve wrench on
calibration gas cylinder to original position (turn clockwise 90°)
·
Click on Save to
confirm updated gas calibration parameters
Gas calibration for the Parvo Medics consolidated system (perform prior to each test)
·
From main menu, click on
Gas Calibration
·
On the gas calibration
configuration windows, click OK
·
Wait unti gas readings
shown at the bottom of the window become stable then click OK to start
calibration procedure
·
Disconnect plastic sampling
tubing from the front of the mixing chamber (left side of cart)
·
On the large green gas
cylinder located from the lab door, turn the round high-pressure valve
clockwise until completely open, then back ½ turn
·
Turn on high-pressure
rod valve just until able to hear gas flow from end of plastic tubing
·
Stick end of plastic
tubing from calibration cylinder over smaller tubing that was previously
attached to mixing chamber-Do
NOT make a tight seal with this connection!
·
Click OK to begin
sample calibration gas. When finished,
return calibration tubing back to gas cylinder and TURN OFF GAS! Reconnect sampling line to mixing chamber
·
Click on Save to
confirm updated gas calibration parameters
Flowmeter calibration (perform only twice daily)
·
Attach one end of the
short (2 ft) blue hose to the 3-L calibration syringe and the other end to the
left port of the mixing chamber at the back of the cart
·
From main menu, click on
Flowmeter Calibration
·
Click on Sample
Baseline
·
When prompted, deliver
10 complete strokes at various flow rates (only the last 5 are used for the
calibration)
·
Click on Save to
confirm flowmeter calibration parameters
Performing metabolic (VO2)
measurements
·
From main menu, click on
Metabolic Measurement
·
Within Patient
Information screen, provide requested information; click OK when
finished
·
Click OK to
automatically sample base-line ambient air; click OK after sample for at
least 10 s
·
When ready to begin
testing, click OK
o
Before beginning test,
properly adjust subject's head gear and nose clip
o
If measuring HR, fit
subject with HR transmitter and attach HR monitor from cart to Velcro strip on
handrail of the treadmill
·
To pause data collection
during a test, click on Pause; click Resume to begin collecting
data again
·
To end the test, click End
Test, and when prompted to save the data, click OK
Viewing / Printing metabolic reports
·
From main menu, click on
Reports
·
In the OUS Report screen,
the top window contains the most recent test and the bottom window contains all
the saved tests
·
Select desired test and
report by clicking appropriate buttons